Transjugular lead fragment extraction to improve ^

tricuspid regurgitation

CrossMark

Robert D. Schaller, DO,* Mouhannad M. Sadek, MD,* Jeffrey J. Luebbert, MD,t Jian-Fang Ren, MD,* Francis E. Marchlinski, MD, FHRS*

From the * Hospital of the University of Pennsylvania, Philadelphia, Pennsylvania, and ^Pennsylvania Hospital, Philadelphia, Pennsylvania.

Introduction

A 28-year-old-man with a history of human immunodeficiency virus (HIV), end-stage renal disease requiring hemodialysis, and HIV-related cardiomyopathy with a severely reduced left ventricular ejection fraction (LVEF) and severe tricuspid regurgitation (TR) per a transthoracic echocardiogram (TTE) underwent implantation of a single-chamber implantable cardioverter-defibrillator (ICD) in 2009 (Riata 1581, St Jude Medical, St Paul, MN). Because of lead failure resulting in electrical noise manifesting as nonphysiological signals, a totally subcutaneous ICD (Boston Scientific, Inc, Natick, MA) was implanted. The patient was referred for percutaneous extraction of the endovascular lead because of his young age, relatively short lead dwell time, high risk of a future blood stream infection in an HIVpositive patient on chronic hemodialysis, and patient preference after an informed discussion of risks and benefits. The procedure was scheduled for 2 months after subcutaneous ICD implantation to facilitate healing of the new device sites.

There was no fluoroscopic evidence of conductor extrusion (Figure 1A). Examination of the pocket revealed severe calcification of the capsule, requiring complete capsulectomy for the removal of the ICD generator (Figure 2). A trans-venous lead extraction procedure was performed using a combination of a locking stylet (LLD EZ, Spectranetics, Colorado Springs, CO) and a 14-F outer diameter SLS II Excimer laser-powered sheath (Spectranetics). Of note, the locking style could not be advanced past the distal coil of the

ABBREVIATIONS HIV = human immunodeficiency virus; ICD = implantable cardioverter-defibrillator; ICE = intracardiac echocardiography; IJ = internal jugular; LVEF = left ventricular ejection fraction; RV = right ventricular; SVC = superior vena cava; TTE = transthoracic echocardiogram; TR = tricuspid regurgitation; TV =

tricuspid valve (Heart Rhythm Case Reports 2015;1:95-98)_

This work was supported in part by the F. Harlan Batrus Research Fund. Address reprint requests and correspondence: Dr Robert Schaller, Hospital of the University of Pennsylvania, 3400 Spruce St, 9 Founders Pavilion, Philadelphia, PA 19104. E-mail address: robert.schaller@uphs. upenn.edu.

right ventricular (RV) lead. Preoperative intracardiac echo-cardiography (ICE) imaging with an 8-F phased-array AcuNav catheter (Acuson Corporation, Mountain View, CA) demonstrated an LVEF of 5%—10% with spontaneous echo contrast in the left ventricle (Figure 3A), severe TR, as well as enhanced echogenicity at the lead-myocardial interface along the inferoseptal wall (Figure 3B), suggesting a potentially strong binding site. Antegrade laser application advanced the sheath to the superior vena cava (SVC) where progress was impeded. Because of the tenacious adhesion at the RV apex, further traction resulted in severing of the lead and the locking stylet in the SVC and the right atrium (Figure 1B) before upsizing to a 16-F outer diameter laser-powered sheath or rotational dilator sheath could be considered. The lead was snared from a femoral approach with a 20-mm Amplatz GooseNeck snare (ev3 Inc, Plymouth, MN), as some of the inferiorly directed frayed conductors were amenable to this approach. Despite reinforcement with a steerable sheath (Agilis, St Jude Medical Inc), repeated attempts to disengage the lead from the myocardium resulted in removal of proximal fragments, with inability to extract the distal portion of the lead (Figure 1C). Each attempt resulted in severe hypotension due to partial obstruction of the tricuspid valve (TV) annulus. This resulted in frayed ICD conductors floating freely in the TV as well as a proximal high-voltage coil fragment affixed to the SVC (Figure 1B). No further attempts were made at this point to discuss the risks and benefits of repeated attempts with the patient.

Postoperative ICE imaging revealed normal TV leaflet morphology but worsening of TR, from severe to torrential, on the basis of visual estimation of a dedicated echocardiog-rapher (Figure 3D), likely secondary to the retained lead fragment and frayed conductors restricting valve closure. After a discussion with the patient, a decision was made for a repeat attempt at percutaneous extraction, and this time via a right internal jugular (IJ) approach. Preoperative ICE imaging once again revealed torrential TR. Ultrasound-guided percutaneous access to the right IJ was achieved, and a 25mm GooseNeck snare was advanced through a steerable sheath (Agilis) to the right atrium (Figure 1D). Once the lead was captured, the Agilis sheath was advanced over the lead

2214-0271 © 2015 The Authors. Published by Elsevier Inc. This is an open access article under the CC BY-NC-SA license (http://creativecommons.org/licenses/by-nc-sa/3.0/).

http://dx.doi.org/10.1016/j.hrcr.2014.11.003

KEY TEACHING POINTS

• The decision to extract an abandoned lead is complex. One must weigh the risks and benefits of each approach and individualize it to each patient.

• Percutaneous lead extraction from the right internal jugular vein is a nontraditional and uncommon technique. In the setting of failed subclavian or femoral extraction, this strategy appears safe and should be considered.

• Right ventricular pacing and defibrillator leads can cause worsening of tricuspid regurgitation. Extraction of such leads may help valvular function. This should be one of the considerations as part of a comprehensive lead management strategy.

body enveloping some of the conductors but was ultimately unable to advance to the lead tip. With the lead firmly positioned at the tip of the sheath, traction was used to pull the lead back to the TV (Figures 1E and 3E; Online Supplemental Video 1), where it finally broke free of its endocardial binding site. The lead and the sheath were

removed from the body at the same time, followed by manual pressure at the entrance site. Next, the Agilis sheath was placed in the right femoral vein and used to snare and remove the high-voltage coil fragment in a similar manner. ICE imaging revealed a significant decrease in TR immediately after the procedure compared to before the procedure (Figure 1I). The patient was discharged home the following day in stable condition. Follow-up TTE at 5 months revealed TR to be mild, suggesting that much of the baseline TR was likely caused by mechanical interaction with the RV lead before the index procedure. Of note, the LVEF also improved from 5%—10% to 40% (Online Supplemental Videos 2 and 3).

Discussion

There is an increasing demand for percutaneous lead extraction as indications have expanded and tools and technologies have evolved.1 Lead advisories and an increase in device-related infections are some of the reasons why physicians are now more likely to encounter patients with complicated management issues requiring complex and inventive solutions.

One must always balance the risks and benefits of lead extraction vs abandonment on a case-by-case basis. In this

Figure 1 Fluoroscopic images for the extraction of an implantable cardioverter-defibrillator (ICD) lead: (A) baseline image, (B) broken lead with frayed conductors in the tricuspid valve and coil fragment along the superior vena cava, (C) snare attempt with traction from below, (D) lead fragment snared from above, (E) snare attempt with traction from above (note the deformity of the cardiac silhouette before disengagement), and (F) immediately after extraction.

SchaLLer et al TransjuguLar Lead Extraction to Improve TR

Figure 2 Calcified capsule surrounding the front and back of the implantable cardioverter-defibrillator.

Figure 3 Intracardiac echocardiography images for the extraction of an implantable cardioverter-defibrillator (ICD) lead, with the transducer placed in the right ventricle (RV) or right atrium (RA): (A) severe left ventricular (LV) dysfunction with left ventricular ejection fraction < 5% and spontaneous echo contrast (arrows), (B) degenerated/dysfunctional ICD lead with increased echogenicity and inhomogeneity in the RV (arrows), and (C) in the RA (arrows), (D) torrential tricuspid regurgitation (TR) indicated by Doppler color flow imaging before extraction and pulmonary artery systolic pressure estimated at 51 mm Hg, (E) during extraction of the ICD lead through the sheath with obliteration of the tricuspid valve (TV) annulus (arrow), (F) RA and RV images immediately after extraction; (G) Gross specimen of the extracted ICD lead with frayed conductor cables, (H) Gross specimen of the RV lead tip containing myocardial tissue, and (I) severe TR after the extraction of the ICD lead significantly reduced severity as compared to torrential TR in panel D. Ao = aorta; mv = mitral valve; RVOT = right ventricular outflow tract.

particular patient, an initial strategy of lead extraction was supported by the patient's young age and excellent functional status despite low LVEF, relatively short lead dwell time, high risk of future systemic infection in the setting of long-term hemodialysis and HIV, high national and international Riata extraction success rates combined with a low major complication rate,2 and patient preference after an informed discussion. In addition, there is an emerging literature on specific problems related to the Riata family of leads, including a possible increased risk of thrombus formation and endocarditis that may further support lead extraction in this situation.3-5

Some experienced lead extractors have advocated the upfront use of the 16-F outer diameter laser-powered sheath (Spectranetics) in this setting. As there is no definitive data regarding the safety or efficacy of this approach, it has been our strategy to start with the smallest appropriate sheath and upsize as needed. Unfortunately, severing of the lead occurred before this consideration. Whether the use of a 16-F sheath would have avoided this complication is unclear. Another potential option is surgical extraction, which is generally reserved for patients with Gram-positive bacter-emia who have failed percutaneous approach as well as patients with a large atrial lead-related thrombus that may result in clinically significant pulmonary embolism. Because of the improved efficacy and safety of percutaneous methods, this approach is rarely required.

Worsening function of the TV during lead implantation and lead extraction has been well described.6'7 In this particular patient with a history of left- and right-sided heart failure and elevated pulmonary arterial pressure, the benefit of a second extraction attempt outweighed the risks of a periprocedural complication. Interestingly, follow-up TTE revealed TR to be mild, suggesting that the RV lead was likely interfering with valve closure before the first procedure. While resultant TV function and LVEF both improved beyond the patient' s baseline status, this must be interpreted with caution, as this was not a specific indication for the index procedure and the mechanism of this dramatic improvement is not completely clear.

Because of the tenacious adhesion process in this patient despite a relatively short lead dwell time, extraction via a femoral approach failed as an inferiorly directed angle of traction was suboptimal (Figure 1C). Extraction from above in this situation conferred a more favorable and efficient

angle, which, despite the absence of countertraction with an outer sheath secondary to technical challenges, resulted in a successful outcome. Extreme caution is advised when simple traction without countertraction at the binding site is used, which is the traditional approach. Bongiorni et al8 have recently described their 15 years' experience of safely and effectively performing transjugular lead extraction. Despite this, extraction from the IJ vein is a nontraditional and uncommon technique because of operator's unfamiliarity. In the case of a lead tip that is severely bound to the ventricular myocardium requiring percutaneous snaring, this strategy should be considered.

Appendix Supplementary data

Supplementary data associated with this article can be found in the online version athttp://dx.doi.org/10.1016/j.hrcr.2014. 11.003.

References

1. Wilkoff BL, Love CJ, Byrd CL, Bongiorni MG, Caiillo RG, Crossley GH III, Epstein LM, Friedman RA, Kennergren CEH, Mitkowski P, Schaerf RHM, Wazni OM. Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA). Heart Rhythm 2009;6: 1085-1104.

2. Brunner MP, Cronin EM, Jacob J, et al. Transvenous extraction of implantable cardioverter-defibrillator leads under advisory—a comparison of Riata, Sprint Fidelis, and non-recalled implantable cardioverter-defibrillator leads. Heart Rhythm 2013;10:1444-1450.

3. Goyal SK, Ellis CR, Rottman JN, Whalen SP. Lead thrombi associated with externalized cables on Riata ICD leads: a case series. J Cardiovasc Electrophysiol 2013;24:1047-1050.

4. Richardson TD, Kolek MJ, Goyal SK, Shoemaker MB, Lewis AA, Rottman JN, Whalen SP, Ellis CR. Comparative outcomes of transvenous extraction of Sprint Fidelis and Riata defibrillator leads: a single center experience. J Cardiovasc Electrophysiol 2014;25:36-42.

5. Maytin M, Wilkoff BL, Brunner M, et al. Multicenter experience with extraction of the Riata/Riata ST ICD lead. Heart Rhythm 2014;9:1613-1618.

6. Lin G, Nishimura RA, Connolly HM, Dearani JA, Sundt TM III, Hayes DL. Severe symptomatic tricuspid valve regurgitation due to permanent pacemaker or implantable cardioverter-defibrillator leads. J Am Coll Cardiol 2005;45: 1672-1675.

7. Franceschi F, Thuny F, Giorgi R, Sanaa I, Peyrouse E, Assouan X, Prévôt S, Bastard E, Habib G, Deharo J-C. Incidence, risk factors, and outcome of traumatic tricuspid regurgitation after percutaneous ventricular lead removal. J Am Coll Cardiol 2009;53:2168-2174.

8. Bongiorni MG, Segreti L, Di Cori A, Zucchelli G, Viani S, Paperini L, De Lucia R, Boem A, Levorato D, Soldati E. Safety and efficacy of internal transjugular approach for transvenous extraction of implantable cardioverter defibrillator leads. Europace 2014;16:1356-1362.