Journal of Clinical and Experimental Investigations

2014; 5 (3): 362-367 doi: 10.5799/ahinjs.01.2014.03.0420

ORIGINAL ARTICLE / ÖZGÜN ARAÇTIRMA

Evaluation of the analytical performance of Roche immunoturbidimetric immunoglobulin assays and comparing to Dade Behring immunonephelometer

Roche immüntürbidimetrik immunoglobulin analizlerinin analitik performans degerlendirmesi ve Dade Behring immünnefelometri ile kar§ila§tirilmasi

Cevdet Züngün1, Fatma Meriç Yilmaz12, Bagdagül Çakir1

ABSTRACT

Objective: The aim of this study is to evaluate the analytical performance of the Roche immunoturbidimetric immu-noglubulin assays, and also compare the results obtained by Roche immunoturbidimetric method to Dade Behring II Nephelometer.

Methods: Low and high concentration patient samples and internal quality control materials were used to evaluate precision, limit of detection (LoD) linearity, and sample carry over of Roche immunoturbidimetric analyzer. External quality control results were used to evaluate accuracy. For method comparison, 451 patients serum were analyzed using Roche Hitachi Modular P System with an immunoturbidimetric method and Dade Behring BNII immunonephelometer with an immunonephelometric method.

Results: In precision study on Roche immunoturbidi-metric immunoglobulin assays, imprecision values were 1.37%, 3.88% and 5.34% for IgG, A and M, respectively. Correlation coefficents between Roche immunoturbidi-metric analysis and Dade Behring BN II immunoneph-elometer were 0.967, 0.979 and 0.939 for IgG, IgA and IgM respectively.

Conclusion: In immunoturbidimetric Roche Ig analyses precision for IgG was allowable according to the the desirable quality specifications. However, both IgA and IgM had higher imprecisions than the desirable total precision. Immunoturbidimetric Roche Ig analyses were found linear in the reference values but unreliable results were observed in the presence of excess antigen. The same samples performed on Roche analyzers needed to be diluted suggesting Dade Behring had wider linearity range than Roche immunoturbidimetric analyzer in the presence of excess antigen. J Clin Exp Invest 2014; 5 (3): 362-367

Key words: Ig G, Ig A, Ig M, immunoturbidimetry, immu-nonephelometry

Amag: Bu gali§manin amaci Roche immunturbidimetrik immunoglobulin analizlerinin analitik performanslarinin de-gerlendirilmesi yaninda bu metot ile elde edilen sonuglarin immunnefelometrik yontem kullanilan Dade Behring II ci-hazi ile kar§ila§tirilmasidir.

Yontemler: Roche immunturbidimetrik yontemin tekrar-lanabilirlik, alt olgum siniri, linearite ve numune ta§inma etkisi (carry over efekti) degerlendirmelerinde, du§uk ve yuksek Ig konsantrasyonlarina sahip hasta ornekleri ve ig kalite kontrol materyalleri kullanildi. Dogruluk analizi di§ kalite kontrol sonuglari degerlendirilerek yapildi. 451 hasta serumu immunturbidimetrik metot kullanan Roche Hitachi Modular P Sistem ve immunnefelometrik metot kullanan Dade Behring BNII immunnefelometride analiz edilerek metot kar§ila§tirmasi gegekle§tirildi. Bulgular: Roche immunoturbidimetrik immunoglobulin analizlerinin tekrarlanabilirlik gali§malarinda IgG igin ke-sinlik degeri 1.37% iken, IgA ve IgM igin kesinlik degerleri sirasiyla 3.88% ve 5,34% olarak tespit edildi. Roche immunoturbidimetrik ve Dade Behring nefelometrik IgG, A ve M analizleri arasindaki korelasyon katsayilari ise sirasiyla 0.967, 0.979 ve 0.939 olarak bulundu. Sonug: immunoturbidimetrik Roche Ig analizlerinde tekrarlanabilirlik IgA ve M igin kabul edilebilir total presizyon limitlerinin uzerinde bulunurken, IgG igin kabul edilebilir limitler dahilinde bulundu. immunoturbidimetrik Roche Ig analizleri igin alt olgum sinirlarinin ureticinin sagladigi bil-gilerle uyum gosterdigi tespit edildi. immunoturbidimetrik Roche Ig analizlerinin referans araliklari igerisinde lineer oldugu, ancak gok yuksek antikor konsantrasyonlarinda guvenilmez sonuglarin olu§tugu gozlendi. Dade Behring BNII nefelometri analizorunde gali§ilan ayni orneklerin, Roche turbidimetrik metotla gali§ildiginda dilusyona ihtiyag duymasi, Dade Behring II otoanalizorlerinin daha geni§ bir linearite araligina sahip oldugunu ortaya koymu§tur. Anahtar kelimeler: Ig G, Ig A, Ig M, immunoturbidimetri, immunonefelometri

1 Clinical Biochemistry, Ankara Numune Education and Training Hospital Ankara, Turkey

2 Yildirim Beyazit University Faculty of Medicine Biochemistry Department Ankara, Turkey

Correspondence: Fatma Men? Yilmaz, Ankara Numune Education and Training Hospital Clinical Biochemistry, Ankara, Turkey Email: fatmamericyilmaz@hotmail.com

Received: 01.07.2014, Accepted: 25.08.2014 Copyright © JCEI / Journal of Clinical and Experimental Investigations 2014, All rights reserved

INTRODUCTION

Turbidimetry and nephelometry are both based on the scattering of radiation by a solution containing dispersed particulate matter [1]. The main technical difference in nephelometry is that the amount of radiation scattered by the particles is measured by viewing at right angles to the incident light while in turbidimetry the amount of radiation transmitted by the suspension is measured. The fraction of the scattered radiation and the reasons directing the scatters such as the number of particles, their size and shape are the main factors in making the choice between turbidimetry and nephelometry. In the presence of many particles leading a large scale of scattering, turbidimetry generally yields more reliable results. On the other hand nephelometry is preferable at low concentrations of analytes owing to easier measuring of a small amount of scattered radiation than a change in density of transmitted radiation Immunoglobulins (Ig) G, A and M are part of analytes classified as "specific proteins" that have wide ranging clinical utility. Single radial immunodiffusion (RID) and electro-immunodiffusion which is a variation of RID have been early methods of immunoglobulin quantitation used in clinical settings. Ig testing performed using methodology of immunonephelometry or immunoturbidimetry on routine clinical chemistry analyzers providing faster results have replaced RID techniques. Automation of Ig quantitation on generalpurpose clinical chemistry systems provides random access, rapid analysis, high volume testing, and cost reduction [2]. The International Federation of Clinical Chemistry (IFCC) and Laboratory Medicine's Committee on Plasma Proteins has designated both im-munonephelometry and immunoturbidimetry as reference methods for quantitation of IgG, IgA and IgM in serum or plasma [3].

The aim of this study is to evaluate the analytical performance of the Roche immunoturbidimet-ric immunoglubulin G, A and M assays in terms of linearity in the presence of excess antigen, limit of detection (LoD) of the assays, sample carry over, precision and accuracy. Also a method comparison study was performed employing Dade Behring II nephelometer as the reference method for method validation of Roche immunoturbidimetric immuno-globulin assays.

METHODS

1. Evaluation of Precision for Roche immunoturbidimetric Ig assays

The precision studies were performed according to Clinical and Laboratory Standarts Institute Educa-

tion Protocol 5 (CLSI EP-5A2), Evaluation of Precision Performance of Quantitative Measurement Methods [4]. Within-run imprecisions were determined using low and high concentration internal quality control materials. Within-run percentage coefficient variations (CVw %) were based on 20 replicate measurements of selected samples per analytical run. Total precisions (CVa%) were obtained measuring human serum aliquotes in an internal protocol as 1 run per day and 10 days.

2. Limit of detections (LoD) of assays on Roche immunoturbidimetric Ig assays

Twenty replicates of the lowest concentration calibrator material and saline were used to calculate the limit of blank (LoB) and LoD following CLSI EP17-A [5]. Standart deviations (SD) and mean values of saline (blank) were used to calculate LoB from the formula LoB = Meanblank + 1.45 SDblank. LoD was determined using LoB value and 1.645 fold SD of lowest concentration calibrator material mean value ( LoD = LoB + 1.645.SD lowest calibrator level) [5].

3. Assessment of linearity for Roche immunoturbidimetric Ig assays

Linearity across the reference range values

In present study we also evaluated the linearity of Roche immunoturbidimetric IgG, A and M assays across the range of reference intervals. Six dilutions (6 (high): 1 (low), through 1 (high): 6 (low)) of low and high level commercial quality control materials were analyzed. Diluted samples were run in duplicate in a single run on Roche Hitachi Modular P analyzer.

Linearity in the presence of excess antigen

Linearity was evaluated following the directions present in CLSI EP6-A [6]. Six dilutions derived from high and low concentration patient samples for each IgG, IgA and IgM were used to assess linearity in the presence of excess antigen. All three samples were diluted 6 (high): 1 (low), through 1 (high): 6 (low). Diluted samples were run in duplicate in a single run on Roche Hitachi Modular P analyzer. The observed results were compared to the theoretical (calculated) results for each dilutions. Upper limits of linearity for IgG, IgA and IgM were determined following ± 10% linearity recovery specifications between the equal concentration ranges of serial dilutions. The average concentration point that the diluted sample exceeded ± 10% linearity recovery was determined as the upper limit of linearity for related immunoglobulin.

4. Sample carry-over on Roche immunoturbidimetric Ig assays

Three replicates of saline measurements (s1, s2 and s3) immediately after three replicates of high concentration samples of IgG (2403 mg/dL), IgA (612 mg/dL) and IgM (416 mg/dL) assays (a1, a2 and a3 for each analytes) were performed to evaluate sample carry over. Carry over was calculated using the formula (s1-s3) / (a3-s3) [7,8].

5. Evaluation of accuracy for Roche immunoturbidimetric Ig assays

The retrospective results of our laboratory in RIQAS external quality results were observed to evaluate accuracy for the peer group of Roche Hitachi analyzer.

6. Subjects and ethics

451 patient specimens delivered to our laboratory from different clinics of our hospital for routine IgG, A, and M testing were used in the present study. 222 of patients were male aged between 20-88 years (mean age: 60.49 ± 14.13) and 229 of patients were female aged between 19-91 years (mean age: 61.45 ± 14.31). A total of 113 patients (57 male; 56 female) had a diagnosis of multiple myeloma. 93 patients (63 male and 30 female) had renal disorders as acute or chronic renal deficiency and ne-phrotic syndrome and 245 remaining were patients with various diagnoses including diabetes mellitus, anemia, leukocyte disorders, trombocytopenia, viral hepatitis and different malignancies. All procedures in the present study were approved by our institution's responsible committee in accordance with the Helsinki Declaration of 1975.

7. Specimens and analyzers for method comparison study

Fasting blood samples were obtained from patients to plain tubes. Sera were separated after centrifu-gation at 1500 g for 10 minutes and analyzed using Roche Hitachi Modular P System (Roche Diagnostics GmBH Mannheim, Germany) with an immunoturbidimetric method and Dade Behring BNII immu-nonephelometer (Dade Behring Canada Inc.) with an immunonephelometric method.

8. Assessment of total error for Roche immunoturbidimetric Ig assays

The total error (TE) can be calculated in a number of ways. The most usual way is to add bias and

precision linearly [9]. We calculated TE for the immunoturbidimetric method according to the formula below;

TE = Bias % + CV

Absolute value of % average of Bland&Altman plot differences were considered to be % bias and and total CVs of IgG, A and M as stated in Table 1.

Table 1. Total and within-run imprecision values of IgG, A and M

Mean SD CVw% CVa

(mg/dL) (mg/dL) (ww(%) %

Low QC level 753.10 13.46 1.78

IgG High QC level 1148.05 33.78 2.94

Human Serum 1223.90 16.80 1.37

Low QC level 143.31 4.02 2.81

IgA High QC level 222.30 3.79 1.70

Human Serum 251.70 9.77 3.88

Low QC level 75.47 4.46 5.91

IgM High QC level 100.47 3.33 3.23

Human Serum 113.8 6.07 5.34

Low QC: low concentration quality control material, High QC: high concentration quality control material, SD: standart deviation, CVw: within-run precision, CVa: total precision

9. Statistical Analysis

Method comparison results were assessed using Bland&Altman difference plots analysis to calculate the bias and limits of agreement. Data normality was assessed using Kolmogorov - Smirnov test. A Paired t - test or Wilcoxon test was used to compare the mean values of methods. Values of p<0.05 were accepted as statistically significant. The non-parametric test of Spearman's rank and Pearson correlation tests were used to evaluate correlation between the results. MedCalc statistical software (ver.12.3.0.0.; Belgium) and SPSS 13.0 for Windows software (Chicago, IL) were used to perform calculations.

RESULTS

1. Precision of Roche immunoturbidimetric Ig assays

Calculation of within-run percentage coefficient variations (CVw %) were based on 20 replicate measurements of selected samples per analytical run and total CVa % (between-run) values obtained

measuring human serum aliquotes in an internal protocol as 1 run per day and 10 days. The total results of precision studies are represented in Table 1.

2. Limit of Detections (LoD) of Assays for Roche immunoturbidimetric Ig assays

LoD values observed for immunoturbidimetric IgG, A and M on Roche Hitachi Modular P analyzer were 30.5 mg/dL, 4.63 mg/dL and 3.38 mg/dL, respectively.

3. Linearity for Roche immunoturbidimetric Ig assays

Across the reference range values

Six dilutions of low and high level commercial quality control materials with the concentrations of related analytes which were close to the stated lower and upper limits (723-1110 mg/dL for IgG, 141-218 mg/dL and 67.7-94.1 mg/dL for IgM). Contrast to the presence of excess antigen, correlation between measured and calculated IgG (R2 = 0.989) , IgA (R2 = 0.987) and IgM (R2 = 0.970) was high.

Presence of excess antigen

Six dilutions derived from high and low concentration patient samples for each IgG (8935 mg/dL - 961.4 mg/dL), IgA (7470 mg/dL - 72 mg/dL) and IgM (9184 mg/dL - 63.7 mg/dL) were used to assess linearity in the presence of excess antigen. The observed results were compared to the theoretical (calculated) results for each dilutions of IgG, IgA and IgM. Correlations between the calculated and measured values of IgG (R2 = 0.662), A (R2 = 0.707) and M (R2= 0.685) were all low. Upper linearity limits for IgG, IgA and IgM were determined as 3500 mg/ dL, 775 mg/dL and 1800 mg/dL, respectively.

4. Sample carry-over on Roche immunoturbidimetric Ig assays

The mean results of 3 replicate measurements of IgG, A and M assays performed using saline were 1.6 mg/dL, 0 mg/dL and 7.1 mg/dL, respectively.

5. Accuracy of Roche immunoturbidimetric Ig assays

The results of our laboratory in RIQAS external quality assessments for cycle 47 for previous nine months were in acceptable limits for standard deviation index (SDI < 2), total score (TS > 50) and % deviation (DEV % < defined acceptable limits) for our peer group of Roche Hitachi analyzer.

6. Method comparison study

IgG, A and M were measured in 451 serum samples by Roche Hitachi Modular P and Dade Behring BNII nephelometer. The mean ± SD value of IgG and mean and (min-max) values for IgA and IgM are summarized in Table 2. The obtained Spearman's rank correlation coefficients were 0.967, 0.979 and 0.939 for IgG, IgA and IgM, respectively.

Table 2. Comparison of results between two methods Roche Hitachi Modular P and Dade Behring BNII

n Roche Hitachi Dade Behring BNII R2 (Turbidimetric) (Nephelometric)

IgG (mg/dL) 451 1351.86 ±803.56 1375.95±861.53 0.967

IgA (mg/dL) 451 261 (166-367) 232 (148.5-335) 0.979

IgM (mg/dL) 451 100 (61.5-139.5) 79 (48.0- 112.5) 0.939

R2: indicates the correlation coefficients between the methods. Data represents median mean ± SD for IgG assays and median (min. - max.) values for IgA and M with 95 % confidence interval.

Agreement between results obtained by immunoturbidimetric and nephelometric methods are demonstrated in different plots using Bland&Altman. There was no evidence of systematic bias (found as 0.7%) (limits of agreement, -26.0 to 27.4) for IgG (Figure 1). However the absolute values for systematic bias was 12.2% (limits of agreement, - 46.0 to 21.7) for IgA (Figure 2) and 22.2% (limits of agreement, -72.6 to 28.3) for IgM (Figure 3).

- +1.96 SO

°fi>° 0 27,4

^ 0 0 -1.96 SD

0 -26,0

0 0 1 » ■

2000 4000 6000 8000 10000

Mean of IgG Dade Behring BNII and IgG Roche Hitachi Modular P

Figure 1. Bland&Altman plots of the difference between Ig G assays performed on Roche Hitachi Modular P and Dade Behring BNII nephelometer against the mean Ig G values of 451 patients.

o +1.96 SD

sÉ® ° -12,2

?fo -1.96 SD

w -46,0

I 1 I 1 , 1 ,1.1.1.1

0 1000 2000 3000 4000 5000 6000 7000

Mean of IgA Dade Behring BNII and IgA Roche Hitachi Modular P

Figure 2. Bland&Altman plots of the difference between Ig A assays performed on Roche Hitachi Modular P and Dade Behring BNII nephelometer against the mean Ig A values of 451 patients.

Figure 3. Bland&Altman plots of the difference between Ig M assays performed on Roche Hitachi Modular P and Dade Behring BNII nephelometer against the mean Ig M values of 451 patients.

7. Total error of Roche immunoturbidimetric Ig assays

As calculated using the TE formula, TEs of Roche Hitachi Modular P for IgG, A and M turbidimetric assays were 2.07%, 16.08% and 27.5%, respectively.

DISCUSSION

According to the literature we reviewed, there are no published articles for evaluation of the analytical performance of immunoglobulin assays on Roche Hitachi Modular P immunoturbidimetric method.

Desirable specifications for imprecision were obtained by using half of the with-in biological variation values of IgG, A and M as given in http://www. westgard.com/database1.htm. In precision study we found total precision of IgG acceptable for the desirable quality specifications of total precision that is equal to 2.3 % for IgG. However, both IgA and IgM had higher imprecisions than the desirable total precision which is 2.5% for IgA and 3.0 % for IgM.

The judgment of acceptability depends on what amount of analytical error is allowable without affecting or limiting the use and interpretation of individual test results [10]. Total error concept composed of random error and systematic error was also evaluated for our study group data. Allowable total error limits were obtained from the charts summarized in http://www.westgard.com/database1.htm. We calculated TE for IgG as 2.07% where the allowable total error for this analyte was 8.0%. TE for Roche Hitachi Modular P was also in acceptable limits for IgG as total precision was. However, Ig A with 16.08 % and IgM with 27.54% total error percentages both exceeded the allowable total error limits which are defined as 13.4% for IgA and 16.8% for IgM.

LoD values determined for IgG, A and M were 30.46 mg/dL, 4.63 mg/dL and 3.38 mg/dL verifying the data stated in manufacturer's kit inserts as 30mg/dL, 4mg/dL and 5 mg/dL, respectively.

The equation in materials and methods section of this paper used for determining carry over assumes that any differences between each of three low concentration samples should contribute to only to with-in run precision if there is no carry over. Sample carry over coefficients (k) for IgG, IgA and IgM were all far below the CVs of the triple measurements indicating the absence of sample carry over.

Our results in Randox International Quality Assessment Scheme (RIQAS) external quality assessments for immunoturbidimetric immunoglobulin analysis were in acceptable limits for our peer group of Roche Hitachi analyzer, however it must be considered that the consensus mean values of IgG, A and M were all in the reference ranges and comparable to the results of immunonephelometry as the reference method. In order to avoid unreliable results in concentrations of immunoglobulins exceeding the linearity limits, further studies are thought to be needed to determine the linear limits for each immunoglobulin method before adapting immunoturbidimetric analysis in a clinical laboratory.

There are different studies in the literature comparing immunoturbidimetric immunoglobulin assays against Dade Behring BN II immunonephelometer. Other analyzers such as Abbott Architect ci8200 had been compared to Beckman Immage for specific protein measurements by Denham et al [1]. In another study, immunoturbidimetric methods on Abbott Architect ci8200 were compared to Dade Behring BNII nephelometer for specific protein assays [11]. But as we reviewed there are no any reports comparing the results between immunoturbidimet-ric Roche Hitachi Modular P immunoglobulin analyses and Dade Behring BN II immunonephelometer Correlation coefficients between Roche immunotur-bidimetric analysis and Dade Behring BN II immu-nonephelometer were 0.967, 0.979 and 0.939 for IgG, IgA and IgM respectively for 95% confidence interval (Table 2). It was suggested that the high correlation coefficients were associated with the mean values of IgG, A and M all within reference intervals. A few data points represented as outliers were excluded which will lead marked discordance between the methods. Thus, correlation coefficients (R2) might be affected due to these excluded data.

In conclusion, Roche immunoturbidimetric Ig analysis were found linear in the reference values but unreliable results were observed in the presence of excess antigen, because of impaired linearity. We found upper limits of linearity for IgG (3500 mg/dL) and IgA (775 mg/dL) comparable to data presented in Roche IgG (3500mg/dL) and IgA (800 mg/dL) manufacturer's kit inserts. For IgM, upper limit of linearity was approximately 1800 mg/dL was higher than the upper limit of linearity data given in Roche IgM (650 mg/dL) kit insert. In our study, patient samples including excess antigen performed on Dade Behring BN II nephelometer could be measured without dilution. The same samples performed on Roche Hitachi analyzers needed to be diluted suggesting BN II nephelometer had wider linearity range than Roche Hitachi immunoturbidimetric analyzer in the presence of excess antigen. We worked out the problem of linearity, diluting the samples serially which were including excess amounts of immunoglobulins till the values decreased to the midpoint of the reference values and following the ±

10 % linearity recovery specifications between the equal concentration ranges of serial dilutions.

Declaration of interest: The authors report no declarations of interest. The authors alone are responsible for the content and writing of the paper.

REFERENCES

1. Morais IPA, Toth IV, Rangel AOSS. Turbidimetric and

nephelometric flow analysis: Concepts and Applications. Spectroscopy Letters. 2006;39:547-579.

2. Denham E, Mohn B, Tucker L, et al. Evaluation of im-

munoturbidimetric specific protein methods using the Architect ci8200: comparison with immunonephelom-etry.Clin Biochem. 2007;44:529-536.

3. Blirup-Jensen S. Protein standardization III: Method optimization basic principles for quantitative determination of human serum proteins on automated instruments based on turbidimetry or nephelometry. Clin Chem Lab Med 2001;39:1098-1109.

4. NCCLS EP5-A2. Evaluation of precision performance

of quantitave measurement methods; approved guideline. NCCLS 2004;Villanova, Pennsylvania.

5. NCCLS EP17-A. Protocols for determination of limits of

detection and limits of quantitation; approved guideline. NCCLS 2004;Villanova, Pennsylvania.

6. NCCLS EP6-A. Evaluation of the linearity of quantitaive

measurement procedures: A statistical approach; approved guideline. NCCLS 2003;Villanova, Pennsylvania.

7. Chen Y, Yazdanpanah M, Hoffman BR, et al. Rapid de-

termination of serum testosterone by liquid chroma-tography-isotope dilution tandem mass spectrometry and a split sample comparison with three automated immunoassays. Clin Biochem 2009;42:484-490.

8. Chen Y, Yazdanpanah M, Wang XY, et al. Direct mea-

surement of serum free testosterone by ultrafiltration followed by liquid chromatography tandem mass spectrometry. Clin Biochem. 2010;43:490-496.

9. Fraser C, editor. Biological Variation:From Principles to

Practice. Washington,DC: AACC Press 2001;29-67.

10. Ricos C, Alvarez V, Cava F, et al. Current databases on biological variation: pros, cons and progress. Scand J Clin Lab Invest 1999;59:491-500.

11. Mali B, Armbruster D, Serediak E, Ottenbreit T. Comparison of immunoturbidimetric and immunoneph-elometric assays for specific proteins. Clin Biochem 2009;42:1568-1571.

Copyright of Journal of Clinical & Experimental Investigations / Klinik ve Deneysel Arastirmalar Dergisi is the property of Journal of Clinical & Experimental Investigations and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use.